IQCP example: the three parts and a sample risk assessment layout
Short answer
An IQCP has three parts: a risk assessment covering at least the specimen, test system, reagent, environment and testing personnel; a quality control plan built from it; and a quality assessment that monitors it. The laboratory director signs and dates it before it's used.
An IQCP has three parts: a risk assessment, a quality control plan and a quality assessment. This page shows what goes in each, with an outline example of a risk assessment.
Reviewed 13 Nov 2026 against the sources linked below. General information, not regulatory or legal advice.
The three parts
- Risk assessment. It covers at least five components: specimen, test system, reagent, environment and testing personnel, across the preanalytic, analytic and postanalytic phases. It uses the laboratory's own data, in its own setting, along with the manufacturer's information.
- Quality control plan. It states the number, type and frequency of controls and the criteria for accepting them. It must give at least what the manufacturer's instructions require.
- Quality assessment. Ongoing review of control records, proficiency testing, corrective actions and complaints. When a failure is found, the laboratory investigates and updates the risk assessment and the plan.
An example risk assessment, in outline
| Component | Example of what could go wrong | Example of how it is controlled |
|---|---|---|
| Specimen | Hemolyzed or short sample gives a wrong result | Rejection criteria in the procedure; collection training |
| Test system | Cartridge or optics fault not seen by the operator | Built-in electronic checks, as described by the manufacturer; external controls at the frequency the plan sets |
| Reagent | Cartridges damaged by heat in shipping or storage | Temperature check at receipt; storage log; controls with each new lot and shipment |
| Environment | Room outside the instrument's operating range | Room temperature and humidity log against the manual's range |
| Testing personnel | Occasional operator skips a step | Training, competency assessment and limiting who may test |
This is an illustration of the layout for a cartridge-based analyzer. It is not a plan. Your risks, your data and the control frequency are specific to your test system and laboratory, and the frequency is your director's decision.
Common mistakes
- Copying a manufacturer's or another laboratory's IQCP without the laboratory's own data.
- Leaving out one of the five components.
- No director signature and date before the plan was used.
- Reducing external controls with nothing in the risk assessment to support it.
- No quality assessment after the plan started.
Templates
The IQCP Kit has the policy, a risk assessment worksheet covering all five components, the quality control plan, the quality assessment review, a supporting data summary and the director approval page. You can read the IQCP policy in full first.
What is an IQCP?
An Individualized Quality Control Plan is a laboratory's own quality control plan for a test system, built on a risk assessment. It is the CLIA option for a nonwaived test when the laboratory does not follow the default control requirements in 42 CFR 493.1256(d).
What are the three parts of an IQCP?
A risk assessment, a quality control plan, and a quality assessment that monitors whether the plan is working.
Who approves an IQCP?
The laboratory director reviews, signs and dates it before it is put into use, and remains responsible for it.
Can an IQCP set controls below what the manufacturer requires?
No. The plan may not be less stringent than the manufacturer's instructions.
Sources
Summaries on this page are ours and are written for convenience. The regulation and the agency documents control.