Home / Guides / Proficiency testing
CLIA proficiency testing requirements: enrollment, handling, scoring and referral
Watch the video1:52 on YouTube, with captions
Short answer
Enroll in a CMS-approved PT program for every regulated analyte you test at moderate or high complexity, and test PT samples exactly as you test patients, never sending them or their results to another laboratory. A passing score is 80% for most analytes and 100% for ABO group, D typing, antibody detection and compatibility testing.
Every laboratory that performs nonwaived testing of a regulated analyte must enroll in an approved proficiency testing (PT) program, test the samples exactly as it tests patients, and review every result. Unsuccessful PT was the most-cited condition-level CLIA deficiency in CMS's 2024 data, and PT referral is among the few violations that can end a laboratory's certificate outright.
Reviewed 16 Nov 2026 against the sources linked below. General information, not regulatory or legal advice.
Who must enroll
- Every laboratory that performs nonwaived testing of a regulated analyte must enroll in an HHS-approved PT program for that analyte, specialty or subspecialty (42 CFR 493.801).
- The regulated analytes are listed in Subpart I. See the full list with acceptance limits, or the approved PT programs for 2026.
- Certificate of Waiver sites don't need PT. If one enrolls anyway, the referral rules apply.
- Accredited laboratories meet their accreditation organization's PT rules too, which can be stricter in detail.
- For nonwaived tests that aren't regulated analytes, verify accuracy twice a year instead (493.1236(c)).
Handle PT samples like patient specimens
- Test them in the regular workload, by the people who usually run the test, on the primary method, the same number of times as a patient specimen.
- Rotate PT among everyone who performs the test.
- Never discuss results with another laboratory before the program's cut-off date.
- Never send a sample to another laboratory. If you would refer that test, report “would refer” or “test not performed”.
- The analyst and the laboratory director sign the attestation that samples were tested like patient specimens.
- Submit before the deadline. CMS's brochure notes that late results are scored zero.
The free PT event record documents each step from receipt to submission.
Scoring: unsatisfactory and unsuccessful
- Satisfactory: usually 80 percent of samples within limits for each analyte in an event; 100 percent for ABO group, D typing, unexpected antibody detection and compatibility testing.
- Unsatisfactory: missing the satisfactory score for an analyte, subspecialty or specialty in one event.
- Unsuccessful: unsatisfactory for the same analyte, subspecialty or specialty in two consecutive events, or two of three. This is a condition-level deficiency, and the most-cited one in CMS's 2024 list (42 CFR 493.803).
- Review every evaluation, including passing ones and samples marked “not graded”, which score an artificial 100 percent. Document the review (493.1236).
After a failure
- Investigate each unacceptable result: clerical error, sample handling, calibration, reagents, instrument, training, or the PT material itself. Record the cause and the correction.
- Check whether patient results were affected over the same period, and act on any that were.
- After unsuccessful performance, a laboratory can stop testing that analyte voluntarily and tell its State Agency. It can resume after two consecutive successful PT events.
- If CMS orders the laboratory to stop, Medicare and Medicaid payment for the test and the certificate can be limited or suspended for six months.
- For a first failure with no history, CMS may offer technical assistance and training instead of a sanction.
PT referral: the line you can't cross
Sending PT samples to another laboratory is one of the most serious CLIA violations. CMS lists the consequences: revocation, suspension or limitation of the certificate, a ban of up to two years on the director directing and the owner owning or operating a laboratory, a directed plan of correction, civil money penalties, and listing on CMS's laboratory registry.
If your laboratory receives PT samples from another laboratory, don't test them. Notify CMS, your State Agency and any accreditation organization with the sender's name and the tests requested. If a referral has already happened, contact your State Agency or CMS at LabExcellence@cms.hhs.gov.
What changed in July 2024
- CMS's final rule CMS-3355-F took effect for PT on July 11, 2024.
- 29 analytes were added to Subpart I, including hemoglobin A1c (target value ±8%), BNP, proBNP, troponin I and T, PSA and vancomycin. Five were removed, including LDH isoenzymes, quinidine and procainamide.
- Acceptance limits were revised for nearly every analyte, many now percentage-based with a fixed floor.
- Microbiology moved from lists of named services to broad organism categories, with fewer mixed cultures and at least two susceptibility samples per bacteriology event.
Records to keep
Keep every PT record for at least two years from the event date: the signed attestation, worksheets and instrument printouts, a copy of the result form, the evaluation, and the review and any corrective action. The free PT event record covers handling. The Proficiency Testing Kit adds the PT policy, a result review and corrective action form, an enrollment and calendar tracker, and a twice-yearly accuracy verification record for tests without required PT.
Do waived laboratories need proficiency testing?
No. CLIA requires proficiency testing only for nonwaived tests of the regulated analytes. A waived site may enroll voluntarily, but once enrolled, the PT referral rules apply to it in full.
What is a passing proficiency testing score?
For most analytes, 80 percent of the samples in an event must be within the acceptable limits. ABO group, D (Rho) typing, unexpected antibody detection and compatibility testing require 100 percent.
Can we send a PT sample to our other site or reference lab?
No. Never send a PT sample, or part of one, to another laboratory, even one in your own organization and even if you send patient specimens there. If you would normally refer that test, mark it as “would refer” or “test not performed” on the result form.
How long do we keep proficiency testing records?
At least two years from the date of the event, including the signed attestation, worksheets, instrument printouts, the result form and the evaluation.
What about tests that have no required PT?
Verify their accuracy at least twice a year (42 CFR 493.1236(c)), for example by splitting a patient specimen with another laboratory and having the director review the results, or by enrolling in a PT program anyway.
Sources
- CMS, Proficiency testing and PT referral brochure (October 2024, PDF)
- CMS, QSO-24-15-CLIA: PT analytes and acceptable performance final rule (July 2024, PDF)
- Federal Register, CMS-3355-F, CLIA proficiency testing analytes and acceptable performance (11 July 2022)
- CMS, Proficiency testing programs
- CMS: CLIA top ten deficiencies in the nation, 2024 (PDF)
- 42 CFR Part 493, Subparts H and I (eCFR)
Summaries on this page are ours and are written for convenience. The regulation and the agency documents control.