CLIA deficiency D5401: procedure manual available and followed

Short answer

D5401 means a test had no written procedure, the procedure wasn't available where testing is done, or staff didn't follow it. The fix is a current, director-approved procedure for every test, kept at the bench, and records showing staff were trained on it.

D5401 is cited when a test has no written procedure, the procedure is not available where testing is done, or staff do not follow it. It was number 3 on CMS's 2024 list of standard-level deficiencies, cited in 635 laboratories.

Reviewed 13 Nov 2026 against the sources linked below. General information, not regulatory or legal advice.

What the regulation requires

  • Have a written procedure manual for every test, assay and examination the laboratory performs.
  • Make it available to the people who do the testing.
  • See that they follow it.

What surveyors typically find

A test on the menu with no written procedure, or staff doing something different from what the procedure says.

How to close the gap

  1. Compare the test menu with the manual, test by test, and write a procedure for anything missing.
  2. Put the current procedures where testing is done, on paper or on screen.
  3. Watch each test being performed and compare what is done with what is written. Change the procedure or retrain, and record which.
  4. Have the laboratory director approve, sign and date each procedure before it is used, and again when it changes.
  5. Keep each retired procedure, with its dates of first use and discontinuance, for two years after it is discontinued.

If you have already been cited

A plan of correction for D5401 says what was fixed, how you checked whether patient results were affected, what changed so it does not recur, and how you will monitor it. See how to write a plan of correction.

Documents for this requirement:

PM-01 Procedure Manual Policy
PM-03 Procedure Manual Index and Review Log

Can the package insert be the procedure?

The manufacturer's instructions may be used for much of the required content. Anything they do not provide, the laboratory adds. The procedure still has to be approved by the director and available to staff.

Who approves a procedure?

The current laboratory director approves, signs and dates each procedure before use, and each change to it. A new director re-approves the manual.

Sources

Summaries on this page are ours and are written for convenience. The regulation and the agency documents control.