CLIA deficiency D5413: storage and operating conditions
Short answer
D5413 means the laboratory hadn't defined, monitored or documented the conditions its reagents, specimens and instruments need, most often temperature. The fix is a stated range for every refrigerator, freezer, room and instrument, a log with no missing days, and a note of what was done whenever a reading was out of range.
D5413 is cited when a laboratory has not defined, monitored or documented the conditions its reagents, specimens and instruments need, most often temperature. It was number 1 on CMS's 2024 list of standard-level deficiencies, cited in 818 laboratories.
Reviewed 13 Nov 2026 against the sources linked below. General information, not regulatory or legal advice.
What the regulation requires
- Define criteria for the conditions that are essential to storing reagents and specimens properly, to operating each test system accurately and reliably, and to reporting results.
- Make those criteria consistent with the manufacturer's instructions where the manufacturer gives them.
- Monitor and document the conditions. Where they apply, that includes water quality, temperature, humidity, and protection of instruments from fluctuations and interruptions in electrical current.
What surveyors typically find
A missing day on a refrigerator log, a range that does not match the package insert, or no action recorded when a reading was out.
How to close the gap
- List every refrigerator, freezer, incubator, room and instrument that has a stated range.
- Take each range from the current package insert or operator manual and write it on the log. Do not carry a range over from an old log.
- Record a reading each day the unit is in use, with initials.
- When a reading is out of range, write down what was done: what was checked, where materials were moved, and whether anything was discarded.
- Have the supervisor or director review the logs on a schedule and sign the review.
If you have already been cited
A plan of correction for D5413 says what was fixed, how you checked whether patient results were affected, what changed so it does not recur, and how you will monitor it. See how to write a plan of correction.
Documents for this requirement:
PM-08 Storage and Operating Conditions Policy
PM-09 Temperature and Humidity Log
How often must temperatures be recorded?
The section says conditions must be monitored and documented and does not give an interval. Follow the manufacturer's instructions and your own written procedure. Most laboratories record each day of use, and surveyors look for gaps against whatever your procedure says.
Does a missed weekend reading count?
It depends on what your procedure requires. If it says daily, a blank is a gap. A written procedure for days the laboratory is closed, such as a minimum and maximum thermometer read on the next working day, closes that gap on paper as well as in practice.
Sources
Summaries on this page are ours and are written for convenience. The regulation and the agency documents control.