CLIA deficiency D5417: expired or deteriorated materials
Short answer
D5417 means reagents, controls, calibrators or supplies were used past their expiration date or after they had deteriorated. The fix is to remove expired stock everywhere it's kept, label anything prepared in-house with preparation and expiration dates, and check dates before use.
D5417 is cited when reagents, controls, calibrators or supplies are used past their expiration date, or when they have deteriorated or are of substandard quality. It was number 5 on CMS's 2024 list of standard-level deficiencies, cited in 607 laboratories.
Reviewed 13 Nov 2026 against the sources linked below. General information, not regulatory or legal advice.
What the regulation requires
- Do not use reagents, solutions, culture media, control materials, calibration materials or other supplies after their expiration date, or when they have deteriorated or are of substandard quality.
- A companion paragraph, 493.1252(c), requires them to be labeled to show identity, storage requirements, preparation and expiration dates, and other pertinent information.
What surveyors typically find
One expired vial in a drawer, or opened reagents with no open date.
How to close the gap
- Check every bench, drawer, refrigerator and crash cart, and remove anything expired.
- Write the date opened, and the new expiration date where the insert shortens it after opening, on each container.
- Record lot numbers and expiration dates when stock is received and when a new lot goes into use.
- Give one person the job of a scheduled expiration check, and keep the record of it.
- Rotate stock so the oldest is used first.
If you have already been cited
A plan of correction for D5417 says what was fixed, how you checked whether patient results were affected, what changed so it does not recur, and how you will monitor it. See how to write a plan of correction.
Documents for this requirement:
PM-08 Storage and Operating Conditions Policy
WT-05 Kit and Reagent Lot Log
Is one expired vial enough for a citation?
It can be. The requirement is that expired materials are not used, and an expired item found in a testing area is typically treated as evidence that it could be.
What about reagents with no printed expiration date?
Follow the manufacturer's instructions. Where the laboratory prepares a reagent itself, it assigns and labels a preparation date and an expiration date.
Sources
Summaries on this page are ours and are written for convenience. The regulation and the agency documents control.