Citations checked against 42 CFR Part 493 on 16 November 2026Independent publisher. Not affiliated with CMS, CDC or any accreditation organization.
CLIA Templates

Home / Point-of-care testing / Glucose meters

Glucose meter (POC glucose) quality control and CLIA requirements

Watch the video1:30 on YouTube, with captions

Short answer

Run controls at the levels and frequency the meter's package insert sets, and use the meter only on the specimens and patients its labeling allows. Never share fingerstick devices, and use a meter on more than one patient only if it's designed for professional use and cleaned and disinfected after every use.

Most glucose meters used in clinics are waived for fingerstick capillary blood. The rules come from the meter's package insert, and two things trip sites up more than anything else: using the meter outside its labeling, and sharing meters or fingerstick devices between patients.

Reviewed 16 Nov 2026 against the sources linked below. General information, not regulatory or legal advice. Your package insert controls.

What CLIA asks of every waived test

A Certificate of Waiver site has one central duty for every test: follow the manufacturer's instructions, including controls, storage, specimen and timing (42 CFR 493.15(e)(1)). Confirm in the FDA CLIA database that your exact device is waived for the specimen you collect, and keep the current package insert with your procedure. CDC notes that most problems found at waived sites come from not following those instructions.

Stay inside the labeling

  • Check the FDA database for your exact meter and specimen type. A meter can be waived for capillary blood and not for venous, arterial or neonatal samples.
  • Read the limitations section of the insert. Using a meter on patients or specimens the insert excludes is off-label use.
  • CMS has treated off-label use as high complexity testing. Its 2014 letter on meters in critically ill patients was withdrawn to draft in 2015, but the safe course is the same: define which patients your meter may be used on and keep to the insert.
  • Put the insert's limitations in your written procedure and your training.

Quality control

  • Run the control levels the insert specifies. Glucose controls usually come in two levels.
  • Take the acceptable range from the test strip vial or the control insert. It can change with every strip lot, so re-enter it for each new lot.
  • Run controls at every point the insert names. Common ones are a new strip vial or lot, a set interval, a meter that was dropped, and results that don't match the patient.
  • If a control is out of range, don't test patients. Check strip expiration and storage, repeat the control and follow the insert.

Infection control for shared meters

  • CDC: never use a fingerstick device for more than one person. Use single-use lancets that retract permanently.
  • CDC: assign a meter to one person whenever possible. If a meter must be shared, use one designed for professional use, not an over-the-counter meter.
  • CDC: clean and disinfect a shared meter after every use, using the method the manufacturer gives. If the manufacturer gives none, don't share that meter.

Mistakes to avoid

  • Using strips after the open-vial discard date printed on the vial.
  • A meter set to the wrong units (mg/dL or mmol/L).
  • A home-use meter shared between patients.
  • Recording a control result without the range it was compared with.

Records to keep

For each control run: date, kit or strip lot and expiration, control lot and expiration, the result against its expected range, whether it was acceptable, what you did if not, and initials. The free waived test QC log covers this for any test. The Waived Testing Kit has a ready-made "Glucose meter" log with the columns for this test, plus the procedure template, lot log and temperature log.

Keep the temperature log for where kits are stored, too: the temperature log is free.

Do glucose meter controls have to be run every day?

Only if the package insert says so. CLIA requires a waived site to follow the manufacturer's instructions (42 CFR 493.15(e)(1)), so the insert's schedule is the requirement. Some sites choose to run them more often than the insert requires; never less.

Can we use a home glucose meter in the clinic?

Not for more than one patient. CDC advises that a meter used on more than one person should be designed for professional use and cleaned and disinfected after every use, following the manufacturer's method.

Sources

Summaries on this page are ours and are written for convenience. The regulation and the agency documents control.

Other point-of-care tests