Point-of-care hemoglobin testing: CLIA requirements and quality control
Short answer
Single-analyte hemoglobin instruments, copper sulfate and spun microhematocrit are waived by the CLIA regulation itself, so a Certificate of Waiver covers them. Run controls as the analyzer's instructions say, and take care with the fingerstick, where most errors start.
Hemoglobin is one of the few tests waived by the regulation itself: CLIA's list includes single-analyte hemoglobin instruments, copper sulfate and spun microhematocrit. Most errors come from the fingerstick, not the analyzer.
Reviewed 16 Nov 2026 against the sources linked below. General information, not regulatory or legal advice. Your package insert controls.
What CLIA asks of every waived test
A Certificate of Waiver site has one central duty for every test: follow the manufacturer's instructions, including controls, storage, specimen and timing (42 CFR 493.15(e)(1)). Confirm in the FDA CLIA database that your exact device is waived for the specimen you collect, and keep the current package insert with your procedure. CDC notes that most problems found at waived sites come from not following those instructions.
What is waived
- 42 CFR 493.15(c) lists hemoglobin by single-analyte instruments with a direct readout, hemoglobin by copper sulfate, and spun microhematocrit as waived tests.
- Some automated blood count analyzers are on CMS's waived list too. A blood count analyzer that isn't waived needs a higher certificate.
- Use the analyzer only for the specimen types its insert names.
Fingerstick technique
- Puncture just off the center of the fingertip pad of the middle or ring finger, after the alcohol has dried.
- Wipe away the first drops as the insert says, then let a free-flowing drop form. Don't squeeze or milk the finger: tissue fluid dilutes the drop and lowers the result.
- Fill the cuvette in one go and check it for air bubbles.
- Read within the time the insert gives. One widely used system's procedure warns that readings taken more than 10 minutes after filling may be wrong.
- Fingerstick drops vary. A 2015 study found hemoglobin changing by more than 2 g/dL between successive drops in some donors, so repeat a result that doesn't fit the patient before acting on it.
Quality control
- Run liquid controls at the levels and frequency the insert specifies. Some analyzers also run an electronic self-test at start-up; that doesn't replace the liquid controls the insert requires.
- Record the control lot and its acceptable range, and re-enter the range for each new control lot.
- Store cuvettes as the label says, and write the opening date on the vial if the insert limits how long opened cuvettes last.
Results to confirm
- Agree action values with your director, such as very low or very high results that need a provider told right away, and record each notification.
- Confirm results above the analyzer's reportable range, or results that don't fit the patient, with a venous sample sent to a laboratory.
- Use reference ranges your director approves. Anemia thresholds differ by age, sex and pregnancy.
Records to keep
For each control run: date, kit or strip lot and expiration, control lot and expiration, the result against its expected range, whether it was acceptable, what you did if not, and initials. The free waived test QC log has these columns and works for this test. The Waived Testing Kit adds a procedure template to build on the package insert, plus the lot log and training checklist.
Keep the temperature log for where kits are stored, too: the temperature log is free.
Is point-of-care hemoglobin a waived test?
Yes, for single-analyte instruments that meet 42 CFR 493.15(c), and for copper sulfate and spun microhematocrit methods. Check your analyzer in the FDA database and follow its insert.
Why do our fingerstick hemoglobins run lower than the laboratory's?
Squeezing the finger mixes tissue fluid into the drop and dilutes it, and successive fingerstick drops vary. Review the technique, and compare with a venous sample when results don't fit.
Sources
- CDC, Ready? Set? Test! Patient testing is important (2024, PDF)
- FDA CLIA test categorization database
- 42 CFR 493.15, Laboratories performing waived tests (eCFR)
- Bond and Richards-Kortum, Drop-to-drop variation in the cellular components of fingerprick blood, Am J Clin Pathol 2015
- Johns Hopkins Pathology, POCT hemoglobin procedure (PDF)
- CMS, Tests granted waived status under CLIA (Transmittal 12415, December 2023, PDF)
Summaries on this page are ours and are written for convenience. The regulation and the agency documents control.