Point-of-care hemoglobin A1c (POC HbA1c): CLIA requirements and quality control

Short answer

Waived point-of-care A1c analyzers are suited to monitoring patients with diabetes. For screening or diagnosis, the 2023 ADA and ADLM guideline limits point-of-care A1c to FDA-approved devices in laboratories certified for moderate complexity testing or higher.

Most point-of-care A1c analyzers sold in the United States are CLIA-waived, and they work well for monitoring patients with diabetes. Using them to screen for or diagnose diabetes is a different matter under current guidelines.

Reviewed 16 Nov 2026 against the sources linked below. General information, not regulatory or legal advice. Your package insert controls.

What CLIA asks of every waived test

A Certificate of Waiver site has one central duty for every test: follow the manufacturer's instructions, including controls, storage, specimen and timing (42 CFR 493.15(e)(1)). Confirm in the FDA CLIA database that your exact device is waived for the specimen you collect, and keep the current package insert with your procedure. CDC notes that most problems found at waived sites come from not following those instructions.

Quality control

  • Run both control levels at the points the insert names, commonly each new lot or shipment, each new tester and at a set interval.
  • Enter the acceptable range from the control insert for the lot in use. Ranges change between lots.
  • Keep controls and test cartridges within the storage and room-temperature times the insert gives.

Monitoring, not diagnosis

  • The 2023 ADA and ADLM laboratory guideline recommends restricting point-of-care A1c for screening and diagnosis to FDA-approved devices in CLIA-certified laboratories that perform moderate complexity testing or higher.
  • A site with only a Certificate of Waiver should use point-of-care A1c to monitor known diabetes and send diagnostic A1c to a laboratory.
  • For diagnosis, the ADA calls for an NGSP-certified method. Check the manufacturer's NGSP certification for any device you use.

Results to question

  • The ADA advises considering interference when an A1c result is above 15%, or when it disagrees with the patient's glucose levels.
  • Hemoglobin variants and conditions that change red cell lifespan can affect A1c. Check the insert's limitations section for the ones that affect your device.
  • Waived A1c devices don't require proficiency testing, but accuracy-based programs are available if you want an outside check.

Records to keep

For each control run: date, kit or strip lot and expiration, control lot and expiration, the result against its expected range, whether it was acceptable, what you did if not, and initials. The free waived test QC log covers this for any test. The Waived Testing Kit has a ready-made "Hemoglobin A1c" log with the columns for this test, plus the procedure template, lot log and temperature log.

Keep the temperature log for where kits are stored, too: the temperature log is free.

Can we diagnose diabetes with our point-of-care A1c?

Not under current guidelines at a waived site. The 2023 ADA and ADLM guideline limits point-of-care A1c for diagnosis to FDA-approved devices in laboratories certified for moderate complexity testing or higher. Use it for monitoring.

Is proficiency testing required for point-of-care A1c?

Not for a waived device. A laboratory using a nonwaived A1c method must enroll in proficiency testing for it.