Home / Point-of-care testing / Rapid strep A
Rapid strep test quality control and CLIA requirements
Short answer
Check the built-in control line on every test and run external positive and negative controls when the package insert says to. Back up a negative result in a child or adolescent with a throat culture.
Rapid strep A antigen tests are among the most common waived tests in physician offices and urgent care. CDC notes they are very specific but less sensitive than a throat culture, which is why a negative result in a child often needs a backup culture.
Reviewed 16 Nov 2026 against the sources linked below. General information, not regulatory or legal advice. Your package insert controls.
What CLIA asks of every waived test
A Certificate of Waiver site has one central duty for every test: follow the manufacturer's instructions, including controls, storage, specimen and timing (42 CFR 493.15(e)(1)). Confirm in the FDA CLIA database that your exact device is waived for the specimen you collect, and keep the current package insert with your procedure. CDC notes that most problems found at waived sites come from not following those instructions.
Quality control
- Every test has a built-in procedural control. If it doesn't appear, the result is invalid: repeat with a new device.
- Run the external positive and negative controls when the insert says to. CDC's waived testing guidance lists each new shipment, each new lot and each new tester as the usual points, and the frequency can't be less than the insert's.
- Record the reason each control run was done, so a reviewer can see the schedule was followed.
Specimen and timing
- Use the swab type the kit specifies. Some kits are validated with particular swabs only.
- Let a refrigerated kit reach room temperature before testing.
- Read at the time the insert gives. CDC warns that reading too early can give invalid or false negative results, and reading late can give false results either way.
Negative results in children
- CDC: rapid tests have high specificity for group A strep but varying sensitivity compared with throat culture.
- CDC: follow up a negative rapid test in children and adolescents older than 3 with a throat culture, and have a way to reach the family if the culture is positive.
- CDC: a backup culture after a negative rapid test isn't routinely needed in adults.
- A throat culture is not a waived test. Arrange it with a reference laboratory and say so in your procedure.
Records to keep
For each control run: date, kit or strip lot and expiration, control lot and expiration, the result against its expected range, whether it was acceptable, what you did if not, and initials. The free waived test QC log covers this for any test. The Waived Testing Kit has a ready-made "Rapid strep A" log with the columns for this test, plus the procedure template, lot log and temperature log.
Keep the temperature log for where kits are stored, too: the temperature log is free.
How often do we run external controls on a rapid strep test?
As often as the package insert says, and never less. CDC's waived testing guidance adds the common practice of running them with each new shipment, each new lot and each new person doing the test.
Is a positive rapid strep result confirmed by culture?
CDC says a positive rapid test or culture confirms group A strep pharyngitis, so a positive doesn't need confirmation. The backup culture is for negative results in children and adolescents.
Sources
- CDC, Ready? Set? Test! Patient testing is important (2024, PDF)
- FDA CLIA test categorization database
- CDC, Clinical guidance for group A streptococcal pharyngitis
Summaries on this page are ours and are written for convenience. The regulation and the agency documents control.