Urine dipstick quality control and CLIA requirements

Short answer

Read each pad at the time the package insert sets, keep strips in the original capped bottle, and run controls as the insert directs. The dipstick is waived, but urine sediment microscopy is not: it needs a PPM certificate or higher.

Urine reagent strips, read by eye or by a waived reader, are on almost every office test menu. The common problems are timing and storage, and one boundary catches sites out: the dipstick is waived, but looking at the urine sediment under a microscope is not.

Reviewed 16 Nov 2026 against the sources linked below. General information, not regulatory or legal advice. Your package insert controls.

What CLIA asks of every waived test

A Certificate of Waiver site has one central duty for every test: follow the manufacturer's instructions, including controls, storage, specimen and timing (42 CFR 493.15(e)(1)). Confirm in the FDA CLIA database that your exact device is waived for the specimen you collect, and keep the current package insert with your procedure. CDC notes that most problems found at waived sites come from not following those instructions.

Quality control

  • Run the control levels the insert specifies, usually a normal and an abnormal level.
  • Compare every pad with the range on the control insert, not just the ones you report most often.
  • Run controls at each point the insert names, commonly each new bottle or lot and each new tester, and when results look wrong.
  • If you use a reader, follow its own schedule for calibration strips or checks as well.

Reading and storage

  • CDC: reading too early can cause invalid or false negative results; reading late can cause false positives from over-development or false negatives from fading.
  • CDC: don't lay strips on a wet or freshly cleaned surface, which can damage the pads.
  • Keep strips in the original bottle with its desiccant, cap closed, and use them by the date the insert gives.
  • Let refrigerated urine reach room temperature if the insert says to.

Where the waiver ends

  • Urine sediment microscopy is a provider-performed microscopy (PPM) procedure under 42 CFR 493.19, not a waived test.
  • It may be done under a PPM certificate only by a physician, midlevel practitioner or dentist during the patient's visit. Anyone else doing it needs a Compliance or Accreditation certificate.
  • The certificate finder checks which certificate your menu needs.

Records to keep

For each control run: date, kit or strip lot and expiration, control lot and expiration, the result against its expected range, whether it was acceptable, what you did if not, and initials. The free waived test QC log covers this for any test. The Waived Testing Kit has a ready-made "Urine dipstick" log with the columns for this test, plus the procedure template, lot log and temperature log.

Keep the temperature log for where kits are stored, too: the temperature log is free.

Is a urine dipstick a waived test?

Most urine reagent strips, read by eye or with many automated readers, are waived. Check your exact strip and reader in the FDA database.

Can a nurse do urine microscopy in our office?

Not under a Certificate of Waiver or a PPM certificate. PPM procedures must be performed by a physician, midlevel practitioner or dentist. Microscopy by anyone else is moderate complexity testing.