Citations checked against 42 CFR Part 493 on 16 November 2026Independent publisher. Not affiliated with CMS, CDC or any accreditation organization.
CLIA Templates

Home / Point-of-care testing / Urine hCG (pregnancy)

Urine pregnancy test (hCG) quality control and CLIA requirements

Short answer

Run controls and read the result at the time the package insert sets. Urine tests detect less hCG than serum tests, so your procedure should say when a negative is repeated or sent for a serum test.

Qualitative urine hCG tests are simple to run and easy to trust too much. They detect less hCG than a serum test does, and some urine specimens interfere with them, so the procedure should say when a negative is repeated or sent for a serum test.

Reviewed 16 Nov 2026 against the sources linked below. General information, not regulatory or legal advice. Your package insert controls.

What CLIA asks of every waived test

A Certificate of Waiver site has one central duty for every test: follow the manufacturer's instructions, including controls, storage, specimen and timing (42 CFR 493.15(e)(1)). Confirm in the FDA CLIA database that your exact device is waived for the specimen you collect, and keep the current package insert with your procedure. CDC notes that most problems found at waived sites come from not following those instructions.

Quality control

  • Check the built-in control line on every test. No line means an invalid test.
  • Run external positive and negative controls when the insert says, commonly each new lot or shipment and each new tester.
  • Bring refrigerated controls and kits to room temperature first, and don't swap caps between control vials.

Why a negative isn't always negative

  • Qualitative devices commonly detect hCG from about 20 or 25 IU/L; quantitative serum assays detect about 5 IU/L. A very early pregnancy can test negative in urine.
  • High levels of a urinary hCG fragment (hCG beta core fragment) can cause false negatives. In one published screening study, levels seen in normal pregnancy affected 9 of 11 devices.
  • Very high intact hCG can saturate the test (the hook effect). Levels in normal pregnancy aren't high enough; it is a concern in conditions that produce extreme hCG.
  • Diluting the urine can turn a falsely negative result positive. Your procedure should name when to repeat, dilute or send a serum test.

Mistakes to avoid

  • Reading after the window in the insert, when evaporation lines can look like a faint positive.
  • Using a specimen type the device isn't labeled for.
  • Recording only "neg" without the kit lot and the control status.

Records to keep

For each control run: date, kit or strip lot and expiration, control lot and expiration, the result against its expected range, whether it was acceptable, what you did if not, and initials. The free waived test QC log covers this for any test. The Waived Testing Kit has a ready-made "Urine hCG" log with the columns for this test, plus the procedure template, lot log and temperature log.

Keep the temperature log for where kits are stored, too: the temperature log is free.

Is a faint line a positive urine hCG result?

Within the reading window, most inserts count any visible test line as positive. Read only at the stated time, and follow your insert's wording exactly.

When should a negative urine pregnancy test be repeated?

When the clinical picture suggests pregnancy anyway, for example very early or with symptoms. A serum quantitative test detects lower levels and isn't affected by the urinary fragment that causes some false negatives.

Sources

Summaries on this page are ours and are written for convenience. The regulation and the agency documents control.

Other point-of-care tests